UL 61010-2-061 PDF

Price range: $36.00 through $72.00

Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-061: Particular Requirements for Laboratory Atomic Spectrometers with Thermal Atomization and Ionization

PDF Format Searchable Printable Bookmarks

1 Scope and object

This clause of Part 1 is applicable except as follows:

Replacement:

Replace the text, except the first paragraph, with the following new text:

This part of IEC 61010 applies to electrically powered laboratory atomic spectrometers with thermal atomization.

NOTE 1 Examples include atomic absorption spectrometers, emission flame photometers, atomic fluorescence spectrophotometers, inductively coupled plasma spectrometers, microwave coupled plasma spectrometers and mass spectrometers, all with thermal atomization and ionization (including tubing and connectors which are provided by the manufacturer for connection to external supplies).

NOTE 2 If all or part of the equipment falls within the scope of one or more other Part 2 documents of IEC 61010 as well as within the scope of this document, consideration is given to those other Part 2 documents.


Summary of Topics

This revision of IEC 61010-2-061, dated September 29, 2023 is being issued to update the title page to reflect the most recent designation as a Reaffirmed American National Standard (ANS). No technical changes have been made.

UL 61010-2-061 is an adoption of IEC 61010-2-061 edition 4, issued September 2018, with no US National Differences.